Specifications - perfectION comb CN
| Measuring range |
8*10-6 – 1*10-2 mol/L 0.2 – 260 mg/L |
| Temperature Range |
0 °C – 80 °C |
| Optimal pH Range |
10 – 14 |
| Connector |
BNC |
| Type of junction |
Click & Clear™ |
| Shaft Material |
Epoxy |
| Sensor Type |
Combined electrode |
| Shaft length |
120 mm |
| Shaft Diameter |
12 mm |
| Cable |
1.2 m |
| Parameter |
ISE |
| Reference electrolyte |
Ion Electrolyte B |
| Type of Membrane |
Solid state |
| Intelligent Sensor Management (ISM®) |
No |
| Temperature Probe |
No |
| Signal type |
Analog |
| Ion |
Cyanide |
| Reference electrolyte Order no. |
51344751 |
| ISA Solution |
10 mol/L NaOH |
| Cable Included |
Yes |
| Connector to Meter |
Flexible (detachable cable) |
| Connector on Sensor |
Fixed 1.2 m cable |
| Ion Type |
Anion |