HORIBA
HORIBA Panel-mount type pH meter with 4-point alarm time sharing proportional control HP-960FTP
HORIBA Panel-mount type pH meter with 4-point alarm time sharing proportional control HP-960FTP
无法加载取货服务可用情况
* Please consult before ordering
Panel-mount type pH meter with 4-point alarm time sharing proportional control
HP-960FTP
A compact water quality meter that offers a wide range of features and functionality.
A pH meter that supports time-proportional control. Suitable for pH monitoring and control in various production processes, as well as wastewater treatment and other applications. We offer a wide range of cleaning devices to accommodate a variety of samples.
Product name | Industrial pH meter | ||
---|---|---|---|
Model | HP-960FTP | ||
Measuring range | 0 pH to 14 pH: 0.01 pH resolution. Temperature 0℃ to 100℃: 1℃ resolution (selectable display). |
||
Repeatability | Within±0.05 pH (Equivalent input) | ||
linearity | Within±0.05 pH (Equivalent input) | ||
Transmission output |
4 mA to 20 mA DC, input/output isolated type |
||
Contact output | Outputs: 4 points Alarm contact output (R1, R2, R3, and R4) Contact type: relay contact, SPDT (1c) Contact rating: 240 V AC 3 A and 30 V DC 3 A (resistance load) Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance. |
||
Contact input | Input: 1 point Contact type: open collector, No-voltage contact Contact function: external input for Hold operation |
||
Control operations |
ON/OFF control
|
||
Calibration function | Two-point automatic calibration or manual calibration. Two-point automatic calibration: with automatic validation of the electrode characteristics. Types of standard solution: pH 2, pH 4, pH 7, pH 9 and pH 10 (JIS) Combination of standard solutions: pH 7 and one of the others Manual calibration: Freely selectable, but the difference should be over 2 pH. Temperature calibration (one point). |
||
Transmission output hold feature |
Selectable from the previous value hold, the preset hold, or the continuous output (However, only the previous value hold is available in the maintenance mode.) | ||
Self-diagnosis function |
Calibration function: Asymmetry potential error, sensitivity error, response speed error and standard solution error. |
||
Temperature detection element |
Selectable from compensation ON (350Ω (25℃), 500Ω (25℃),1 kΩ (0℃)6.8 kΩ (25℃) and 10 kΩ (25℃)) and compensation OFF. |
||
Temperature compensated range |
0℃ to 100℃ | ||
Ambient temperature |
−5℃ to 45℃ | ||
Relative humidity | 20% to 85% (without dew condensation) | ||
Storage temperature |
−25℃ to 65℃ | ||
Power supply | Rated voltage 100 V to 240 V AC, 50 Hz/60 Hz, 10 VA (max.) | ||
Structure | Indoor-use panel installation type Panel case: ABS, Terminal: PBT Panel: IP65 dust and water proof structure |
||
Protective structure | Panel: IP65 (IEC60529, JIS C0920) Rear case: IP20, Terminal: IP00 Class II device (IEC61010-1) Pollution level 2 (IEC61010-1) |
||
Conforming standards |
CE Marking | EMC Directive: EN61326-1 Low Voltage Directive: EN61010-1 RoHS Directive: EN50581 |
Immunity (Industrial electromagnetic environment) Noise increase ±0.05 pH or less |
FCC Rule | FCC Part15 | ||
External dimensions |
96 (W) × 96 (H) × 115 (D) mm Case depth: approx. 105 mm |
||
Mass | Approx. 550 g |
Share