HACH CLT10sc Total Chlorine Analyzer (Panel Only) with Combination pH Sensor PN.LXV45B.99.12022
HACH CLT10sc Total Chlorine Analyzer (Panel Only) with Combination pH Sensor PN.LXV45B.99.12022
无法加载取货服务可用情况
* Please consult before ordering
Hach's answer to reagentless amperometric chlorine measurement.
The CLT10sc Total Chlorine Analyzer measures total chlorine continuously for real time process control. It is based on amperometric technology and is fully compatible with all of Hach's digital controllers. The CLT10sc can be used in most municipal and industrial applications, and is best utilized where waste stream management is a constraint. This analyzer is equipped with a pH sensor for use with the unique Cal Watch advanced diagnostic feature.
Benefits
Exclusive Self Diagnostics
The CLF10 sc and CLT10 sc analyzers leverage Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow.
EPA compliant according to Method 334.0
Chlorine measurement with an amperometric analyzer, such as the CLF10 sc or CLT10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management.
Real-Time Process Control
The CLF10 sc and CLT10 sc analyzers allow for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed.
Compatible with Hach's "Plug and Play" Digital Controllers
The CLF10 sc and CLT10 sc analyzers can be used with any Hach sc digital controller. Whether you're measuring turbidity or chlorine, you only need to learn one controller for all your water analysis measurement points. Hach sc controllers have no complicated wiring or setup procedures. Just plug in any Hach digital sensor and it's ready to use without software configuration.
EPA Compliant According to Method 334.0
In accordance with EPA Method 334.0, the CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required in Method 334.0. (See page 4 for a partial listing of accessories.)
Specifications
Technical Attributes
Accuracy | Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit) |
Accuracy 2 | Free Chlorine: ± of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
Accuracy 3 | Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
Accuracy 4 | Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5 |
Cable Connection | 5 pin, M12 connector |
Cable Length | 1 m (between gateway and sc controller) |
Calibration Method | 1-point or 2-point (zero and slope) calibration |
Certifications | CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Additional certifications when connected to an sc controller |
Connection Drain Line | 1/2 in ID |
Connections | 1/4 in OD |
Controller | Panel only |
Controller Compatibility | sc controller platform |
Description | CLT10sc Total Chlorine Analyzer (Panel Only) with Combination pH Sensor |
Dimensions | Sensor Only: 195 |mm x 25 mm (7.68 in x 0.98 in) |
Fitting Type | USA |
Includes | CL10sc Panel, 1 m digital extension cable, Panel Manual, Combination pH Sensor Manual, Chlorine Sensor Manual. |
Interferences | Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. Total chlorine: chlorine dioxide, ozone and chalk deposits |
Length | Sensor: 195 mm |
Lower Limit of Detection (LOD) | 30 ppb (0.03 ppm) |
Material | Corrosion-resistant materials, fully-submersible (stainless steel, PVC, silicon rubber and polycarbonate) |
Measurement Method | Reagentless, electrochemical, three-electrode amperometric system |
Mounting | Panel or Wall Mounting |
Operating Temperature Range | 0 - 45 °C |
Options | Combination pH Sensor |
Parameter | Total Chlorine |
pH Monitoring Required? | With combination pH sensor |
pH Range | 4 - 9 |
Power Requirements (Voltage) | None |
Pressure Range | 0.5 bar, no pressure impulses and/or vibrations |
Range | 0 - 20 ppm Cl₂ |
Repeatability | 30 ppb or 3%, whichever is greater |
Response Time | Total Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH |
Response Time | Free Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH |
Sample Flow Rate | 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour) |
Sample Temperature | 5 - 45°C (41 - 113°F) |
Storage Conditions | -20 °C to 60 °C dry |
Warranty | 12 months |
Weight | Approximately 5.5 kg (12 lbs) Panel and empty panel-mounted components only |
What's included? | CL10 sc Panel, 1 m digital extension cable, Panel Manual, Combination pH Sensor Manual, Chlorine Sensor Manual. |