HACH
HACH AN-ISE sc Low cost ISE combination Ammonium and Nitrate probe (immersion) without RFID, 10 m cable PN.LXV440.99.00012
HACH AN-ISE sc Low cost ISE combination Ammonium and Nitrate probe (immersion) without RFID, 10 m cable PN.LXV440.99.00012
无法加载取货服务可用情况
* Please consult before ordering
AN-ISE sc: Combination sensor for ammonium and nitrate
The special feature of the probe is the Cartrical technology. This provides reliable measured values and considerably reduced maintenance time and costs in comparison with conventional ISE probes.
Comes as a complete system from a single supplier. The probe is installed on the side of the tank. An SC controller completes the system. It can be extended to include an automatic cleaning unit with a compressor.
Prognosys is a predictive diagnostic system that allows you to be proactive in your maintenance, by alerting you to upcoming instrument issues. Know with confidence whether changes in your measurements are due to changes in your instrument or your water.
This instrument connects to Claros, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations. To unlock the full potential of Claros, insist on Claros Enabled instruments.
Benefits
Trustable results for ammonium and nitrate
Looking for new ways to stretch money and reduce consumable cost? We have the perfect solution for you! Our new combination sensor with ion-selective electrode (ISE) technology was developed to measure both NH4 and NO3 at the same spot. What is unique? The most significant interferences of K and Cl are simultaneously compensated by integrated ISE electrodes, ensuring accurate results 24h/7days.
Easy handling with Cartrical cartridge plus
The compact Cartrical cartridge contains four electrodes which are calibrated to each other before leaving the factory. Through the newly developed Cartrical plus technology, the factory calibration is automatically identified after replacing the cartridge. The compact design does away with tiresome handling and adjustment of individual electrodes: just unscrew the old cartridge, plug in the new one and the sensor is immediately ready to measure.
Simple and intuitive operation
With a menu-guided operation and step-by-step guide with clear text, the combination AN-ISE sc sensor is very easy to use. It works directly with every SC controller without complicated wiring or set up procedures: just plug and play. The common SC controller platform allows intuitive usage not only for ammonium and nitrate but also for other parameters like pH, DO, etc.
Always under control
In combination with SC1000 controller alert messages from AN-ISE sc sensor can be transmitted to the control room or by SMS / e-mail to a mobile phone. By this you are always informed about the status of your processes, early enough to react quickly.
Rely on Hach's application know how
You can trust our ISE experience gathered in many waste water applications around the world. Based on this we have developed the next ISE generation to fulfill your requirements even better. Our application team is always close by.
Specifications
Technical Attributes
Accuracy | 5% of measured value +0.2 mg/L (with standard solutions) NH4-N and NO3-N |
Cable Length | Standard: 10 m (33.8 ft) Extension cables are available as an option in the following lengths: 5, 10, 15, 20, 30, 50 m (16.4, 33.8, 49.2, 65.6, 98.4, 164 ft) |
Calibration Method | Factory calibrated sensor cartridge (CARTRICAL), 1 and 2 point matrix correction |
Controller Compatibility | SC200, SC1000, SC4500. All controllers sold separately |
Data Transfer | Current output, relays and bus communication via SC controller |
Dimensions (D x L) | 84.5 mm x 320 mm |
Flow | < 4 m/s max. velocity |
Includes | Ammonium & Nitrate probe, sensor cartridge, user manual |
Installation Angle | 45 ° ± 15 ° (vertical in flow direction) |
Length | 320 mm |
Lower Limit of Detection (LOD) | 0.2 mg/L NH4–N and NO3-N |
Material | Cartridge: stainless steel (1.4571), PVC, POM, ABS, NBR Cleaning unit: TPE, PUR, stainless steel (1.4571) Sensor: stainless steel (1.4571), ASA + PC, silicon, PVC and PU |
Measurement Method | Potentiometric ion-selective electrodes for ammonium, potassium, nitrate and chloride, reference system and temperature sensor |
Measuring Interval | Continuous |
Measuring uncertainty | ± 5% + 0.2 mg/L (with standard solution) NH4-N and NO3-N |
Operating Temperature Range | Air: -20 - 45 °C (-4 - 113 °F) |
Parameter | Ammonia, Nitrate |
pH Range | 5 - 9 pH |
Process Connection | 1 inch thread |
Protection Class | IP68 |
Range | 0 - 1000 mg/L NH4-N 0 - 1000 mg/L NO3-N 0 - 1000 mg/L K+ 0 - 1000 mg/L Cl- |
Region | US |
Repeatability | ± 5 % of the measured value (with standard solution)+ 0.2 mg/L NH4-N and NO3-N |
Response Time | < 3 min |
Sample Pressure | 0.3 bar max. |
Sample Temperature | 2 - 40°C (35.6 - 104°F) |
Sensor Cartridge | Automatic import of factory calibration data, typical lifetime 6 months |
Sensor Immersion Depth | 0.3 - 3.0 m max. 0.3 bar (4.4 psi) pressure |
Storage Conditions | 5 to 40 °C (41 to 104 °F) sensor cartridge |
Warranty | 12 months |
Weight | 2.38 kg (5.25 lb) |
What's included? | Ammonium & Nitrate probe, sensor cartridge, user manual |
Share